The AccuSign® MET test uses solid-phase immunoassay technol-
ogy for the qualitative detection of methamphetamine in humanurine. The test is based on the principle of the highly specificimmunochemical reactions between antigens and antibodies whichare used for the analysis of specific substances in biological fluids.
The test relies on the competition between the drug conjugates andthe drugs which may be present in the urine sample, for binding toantibodies. In the test procedure, a sample of urine is placed in the
Sample well of the device and is allowed to migrate upward. If thedrug is present in the urine sample, it competes with the drug
conjugate bound to the dye, for the limited antibodies immobilized
on the membrane. If the level of the drug or its metabolite is above
the cutoff level, the drug will saturate the antibodies, thus inhibit-ing the binding of the dye coated with drug conjugates to theantibodies on the membrane. This prevents the formation of a lineon the membrane. Therefore, a drug-positive urine sample will notgenerate a line at the specific drug position in the Result window,indicating a positive result from positive drug competition. A
negative urine sample will generate a line at the specific drug
position in the Result window, indicating a negative result from anabsence of competition with free drugs.
In addition to the Test line(s) that may appear in the Result window,a Control line is present to confirm the viability of the test. This
The AccuSign® MET test is a simple, one-step, immunochromato-
Control line (validation line) should always appear if the test is
graphic assay for the rapid, qualitative detection of methamphet-
conducted properly. Polyclonal sheep anti-mouse IgG antibody is
amine in urine above a cutoff concentration of 1000 ng/mL.
immobilized on the control line. The monoclonal antibody-dye
The AccuSign® MET test provides only a preliminary analytical
conjugates that pass the line will be captured and produce a colored
result. A more specific alternative chemical method must be used
line at the Control position (C). This works as a procedural control,
in order to obtain a confirmed analytical result. Gas chromatog-
confirming that proper sample volume was used and the reagent
raphy, mass spectrometry (GC/MS) is the preferred confirmatory
system at the Control line and the conjugate-color indicator worked
method. Other chemical confirmatory methods are available.
properly. If insufficient sample volume is used, there may not be
Clinical consideration and professional judgment should be
a Control line, indicating the test is invalid. applied to any drug of abuse test result, particularly when preliminary positive results are used.1
The AccuSign® MET test kit contains all the reagents necessary to
Methamphetamine is a potent sympathomimetic agent with thera-
AccuSign® MET device. The test device contains a membrane
peutic applications. The drug can be taken orally, injected, or
strip coated with anti-methamphetamine antibody and a dye
inhaled. Acute higher doses lead to enhanced stimulation of the
pad containing colloidal gold coated with drug-protein (from
central nervous system and induce euphoria, alertness, reduced
purified bovine protein source) conjugate.
appetite, and a sense of increased energy and power.2 Cardiovas-
cular responses to methamphetamine include increased blood
pressure and cardiac arrhythmias. More acute responses includeanxiety, paranoia, hallucinations, psychotic behavior, and eventu-
For in vitro diagnostic use only.
The effects of methamphetamine generally last 2–4 hours, and thedrug has a half-life of 9–24 hours in the body. Methamphetamine
Avoid cross contamination of urine samples by using a new
is excreted in the urine primarily as amphetamine and oxidized and
urine specimen container and dropper for each urine sample.
deaminated derivatives.2 However, 10–20% of methamphetamine
The test kit does not contain any HIV or hepatitis infective
is excreted unchanged. Thus, the presence of the parent compound
in the urine indicates methamphetamine use. Methamphetamine is
Urine specimens are potentially infectious. Proper handling
generally detectable in the urine for 3–5 days, depending on urine
and disposal methods should be established according to good
CONTROL (VALIDATION) LINE (C). The Control/Validation line indicates:1.
If the procedure was followed properly.
If no control line appears, the test is NOT VALID. Repeat the test using a new device, and follow the
Negative = Control line and Specific Drug line
Positive = Control line only; No Specific Drug line
The AccuSign® device should remain in its original sealed
pouch until ready for use. Do not use the test if the pouch isdamaged or the seal is broken. Negative: The appearance of a reddish-purple Control line (C) and a line next to Test position (T) indicates a negative test result;
Do not use the test kit after the expiration date.
i.e., no drug above the cutoff level has been detected. The colorintensities of the Control line and a specific drug line may not be
equal. Any faint line in the Result window, visible in 10 minutes,
The AccuSign® MET test kit should be stored at 2–30°C (35– should be interpreted as negative. A negative test result does not
86°F) in the original sealed pouch. The expiration dating was
indicate the absence of drug in the sample; it only indicates the
established under these storage conditions. sample does not contain drug above the cutoff level in qualitativeterms.Positive: The appearance of a reddish-purple Control line and no
Approximately 110 µL of urine sample is required for each test.
distinct line next to T indicates the test result is positive for MET
Fresh urine specimens do not require any special handling or
(i.e., the specimen contains the drug at a concentration above the
pretreatment. Specimens should be collected in a clean glass or
cutoff level). A positive test result does not provide any indication
plastic container. If testing will not be performed immediately,
of the level of intoxication or urinary concentration of the drug in
specimens should be refrigerated (2–8°C) or frozen. Specimens
the sample; it only indicates the sample contains drug above the
should be brought to room temperature before testing. cutoff level in qualitative terms.
Specimens containing a large amount of particulate matter may
Invalid: A distinct Control line (C) should always appear. The
give inconsistent test results. Such specimens should be clarified
test is invalid if no Control line forms at the C position. Such tests
by centrifuging or allowing to settle before testing.
should be repeated with a new AccuSign® MET test device.
The test procedure consists of adding the urine sample to the
The test is designed for use with unadulterated urine only.
Sample well of the device and watching for the appearance of
There is a possibility that factors such as technical or proce-
colored lines in the Test and Control positions.
dural errors, as well as other substances in the urine samplewhich are not listed in Table 5 below, may interfere with the
1. For each test, open one AccuSign® MET
Adulterants, such as bleach and/or alum, in urine specimens
and label the device with the patient ID.
may produce erroneous results regardless of the method ofanalysis. If adulteration is suspected, the test should be re-
2. Holding the dropper vertically, dispense 3 full
drops (110 µL) of the urine sample into the
The test result read after 10 minutes may not be consistent with
3. Read the result after 3 minutes, but within 10
the original reading obtained within the 10 minute reading
In a separate study, AccuSign® MET was evaluated against speci-
period. The test must be read within 10 minutes of sample
mens confirmed as positive by GC/MS. Of 89 samples confirmed
as positive, 88 samples were positive when tested with AccuSign® MET (98.9% agreement). (Table 2.) Table 2. Accuracy: Comparison of AccuSign® MET with Internal Control: Each AccuSign® MET test device has a built- GC/MS Assay
in control. The Control line is an internal positive procedural
AccuSign®
control. A distinct reddish-purple Control line should appear in theControl position, if the test procedure is performed properly, an
adequate sample volume is used, the sample and reagent are
wicking on the membrane, and the test reagents at the control lineand the conjugate-color indicator are reactive. In addition, if the
test is performed correctly and the device is working properly, thebackground in the Result window will become clear and provide
The precision of AccuSign® MET was determined by carrying out
a distinct result. This may be considered an internal negative
the test with serially diluted standard drug solutions. About 98% of
the samples containing methamphetamine concentrations 25%over the cutoff level consistently showed positive results.
The positive and negative procedural controls contained in each AccuSign® MET test device satisfy the requirements of testing a
The study also included over 40 samples containing ± 25% of the
positive control and a negative control on a daily basis. If the
cutoff level as a challenge of cutoff precision. These results were
Control line does not appear in the Control position, the test is
found to be consistently in agreement with expected test results.
invalid and a new test should be performed. If the problem persists,
Distribution of Random Error:
contact PBM for technical assistance.
Twenty blind samples prepared by spiking various concentrations
External Control: External controls may also be used to assure
of methamphetamine were separately tested by two operators. The
that the reagents are working properly and that the assay procedure
test results from the two operators showed complete agreement.
is followed correctly. It is recommended that a control be tested atregular intervals as good laboratory testing practice. For informa-
tion on how to obtain controls, contact PBM’s Technical Services.
The reproducibility of the test results of AccuSign® MET was
examined at three different sites using a total of 15 blind controls,consisting of 5 negative samples, 5 moderately positive samples
AccuSign® MET is a qualitative assay. The amount of metham-
(containing the drug at a concentration 1.5-2 times the cutoff level),
phetamine and/or their metabolites present in urine cannot be
and 5 strongly positive samples (containing the drug at a concen-
estimated by the assay. The assay results distinguish positive from
tration 4-5 times the cutoff level). The results obtained at these
negative samples. Positive results indicate the samples contain
three sites with these controls demonstrated 100% agreement with
methamphetamine and/or their metabolites above the cutoff con-
The following table lists compounds that are detected by the
The AccuSign® MET test has been shown to detect an average AccuSign® MET test. The specificity of the AccuSign® MET test
cutoff of 1000 ng/mL of D-methamphetamine in urine.
was determined by adding the drugs and drug metabolites listed to
The accuracy of AccuSign® MET was evaluated in comparison to
drug-negative urine specimens and testing with the AccuSign®
a commercially available immunoassay, Syva® EMIT® II. A total
MET test kit. The results are expressed in terms of the concentration
of 320 samples was tested with both procedures. Overall agree-
required to produce a positive result. (Table 3.)
ment of 91.3% was observed, as shown below. (Table 1.)
Table 3. Specificity Table 1. Accuracy: Comparison of AccuSign® MET with Concentration Syva® EMIT® II Compound
The following compounds show no cross-reactivity when tested
with AccuSign® MET at a concentration of 100 µg/mL. (Table 4.) Table 4. Non Cross-Reacting Compounds
Hawks RL, Chiang CN, eds. Urine Testing for Drugs of Abuse.
Rockville, MD: National Institute for Drug Abuse (NIDA), Research
Baselt RC. Disposition of Toxic Drugs and Chemicals in Man. 2nd
Ed., Davis, CA: Biomedical Publ.;1982; p.488. In Vitro Diagnostic Medical Device
“Use By” date in year-month-day format
AccuSign® is a Registered Trademark of Princeton BioMeditech Corporation. MT Promedt Consulting GmbH Princeton BioMeditech Corporation
Julie Bruno, AAHPM Director of Education, started the morning plenary session with a reminder that “life happens” at preconference workshops just as it does in real life, so one of the session speakers had to attend to a family health matter. The schedule was therefore rearranged a bit to provide the two remaining faculty time to cover the additional lectures. The speakers are experts and
REPUBLIQUE DU SENEGAL Un peuple – Un but – Une foi MINISTERE DE L’ECONOMIE ET DES FINANCES DIRECTION DE LA PREVISION ET DE LA STATISTIQUE ENQUETE DE SUIVI DE LA PAUVRETE AU SENEGAL (ESPS_2005) Juin 2005 SOMMAIRE INTRODUCTION……………………………………………………………………………………………….1 I - OBJECTIFS DE L'